Sample preparation

Automatic solid-phase extraction system, SPE100/200/400

SPE100/200/400 series automatic solid-phase extraction system with a highly integrated fluid-line design, a multifunction sample needle and a precise, flexible robotic arm, for faster and more efficient sample preparation.

  • Channels and control: modular main unit; the number of channels can be upgraded on site, in 1/2/4/6/8-channel configurations
  • Continuous processing capacity: up to 52 samples in one continuous run
  • Robotic arm: three-axis X/Y/Z manipulator with linked X/Y axes for efficient operation; an additional H axis moves the slider, combining flexible movement with accurate positioning
  • Sample pump: accurate sampling, resistant to solvent corrosion and high pressure (up to 1 MPa); handles both small and large sample volumes
  • Sample needle: 316L stainless steel, solvent-resistant; the side-bottom opening ensures complete sample loading; the liquid line, gas line and extraction-column sealing are integrated in the needle
  • Remote connection and control: built-in router for connection via Wi-Fi or network cable, for fully automated, unattended operation
  • Puncture sampling: an optional sample needle allows puncture sampling, preventing loss of sample through evaporation
  • Sample and collection volumes: the whole series takes 10–20 mL sample tubes, and 50 mL and larger on up to two channels; the standard collection volume is 10 mL/15 mL, expandable to 50 mL on two channels
  • Pressure sensing: a pressure sensor on every channel as standard gives closed-loop fuzzy control of column pressure, with overpressure alarm and automatic termination of the run
  • Flow rate control: each extraction column has its own sealing cap; a high-pressure seal is made by docking the sample needle and moving the Z-axis pressure block, so the solution passes through the column at the set flow rate, giving good reproducibility and consistency between parallel channels
  • Solvent management: up to 8 reagents, low-solvent warning and liquid-level tracking
  • Soaking and incubation: when the activation or elution solvent is loaded, the liquid level can be held above the sorbent bed for a set time to ensure complete activation or elution
  • Cleaning: several cleaning modes keep the inner and outer walls of the sample needle clean, with no cross-contamination
  • Waste lines: waste liquid is recovered for environmentally sound disposal, and waste of different types is collected separately
  • Extraction columns: compatible with standard 1/3/6 mL columns; other sizes available to order
  • Multi-method extraction: columns of different sizes and different methods run simultaneously without waiting, raising throughput in continuous processing
  • A/B column extraction: several tandem-column extraction modes, with automatic recognition of the A/B column method
  • Online drying/dewatering: nitrogen purge
  • Online concentration: in-situ concentration by nitrogen purge, with programmable concentration time
  • Large-volume loading: continuous, uninterrupted loading of samples over 500 mL; resistant to particle wear and corrosion, with no consumables or routine maintenance required; continuous switching between more than 48 samples
  • Cloud services: upload and download of programs and parameters, online technical support and remote monitoring
  • Scheduled start: once a method is set up, its start time can be scheduled to suit the analyser's timetable
  • Countdown: once a run starts, a countdown makes it easy to plan the end of the run and align it with the analyser in advance
  • Self-check and alarms: after the parameters are set, the software checks them for logic errors and prompts for corrections; the fault alarm system continues or stops the run depending on the fault level

Also listed in: Food technology: Food sample preparation

Specifications

Brand
Hanon · SPE100 +2
Number of channels
1/2/4/X
Independent channel control
Yes
Sample pump
Precision continuous sample pump
Pump flow rate
0.1–120 mL/min
Number of solvents
8; 500 mL solvent bottles
Sample volume
0.1 mL to 20 L
Sample tube volume
10–20 mL; other sizes to order
Collection volume
10 mL/15 mL; other sizes to order
Puncture sampling
Yes
Sample loading
Three-axis X/Y/Z manipulator
Manipulator accuracy
0.01 mm
Sealing technology
Individual sealing caps with docking plug seal
Extraction mode
Positive-pressure extraction
Column pressure
≤80 psi
Pressure sensing
Yes; overpressure alarm, adjustable threshold (5–145 psi)
Column drying
Nitrogen purge
Waste collection
Separate collection, 3 channels
A/B columns in series
Yes (A/B columns)
Cleaning modes
Nitrogen purge, solvent rinse, bubble-agitation cleaning and other modes for the inner and outer walls of the sample needle
Liquid-level tracking
Yes
Solvent level monitoring
Low-solvent warning
Main unit dimensions (H × W × D)
680 × 580 × 670 mm
Power supply
220/110 V, 50/60 Hz
Operating temperature
5–40 °C
Operating humidity
0–90%, non-condensing
Power consumption
60 W/80 W/120 W/<200 W

On request

All options (3)

Variants & pricing

  • Automatic Solid Phase Extraction SPE100

    SPE100On request
  • Automatic Solid Phase Extraction SPE200

    SPE200On request
  • Automatic Solid Phase Extraction SPE400

    SPE400On request