Skip to main content
Saint-Tech

Photo 1 of 6

Fume hoods & safety cabinets

Hazardous material workstation EX-LINE – complete package with fresh air curtain

EN 14175EN 14175-3

Function / construction:

  • Required occupational exposure limits are maintained
  • Ventilation testing in accordance with EN 14175 Part 3 (5.4.4)
  • Maximum user safety
  • Robust construction made of chemically resistant, anodised aluminium profiles
  • Transparent side panes made of 5 mm toughened safety glass
  • Highly effective capture of hazardous substances inside the workstation thanks to special fresh air curtain technology
  • Prepared for connection to the on-site exhaust air system
  • Electrical equipment approved in accordance with 2014/34/EU
  • For regulation-compliant use in potentially explosive areas, Zone 1:
  • Workplace LED lighting: IBExU 16 ATEX 1008, Zone 1, application area II 2 G Ex eb mb op is q IIC T4 Gb
  • Supply air fan: PTB 03 ATEX 3004, Zone 1, application area II 2 G, c T4
  • Switch amplifier: PTB 00 ATEX 2081, application area II (1) GD EEx ia II c
  • Differential pressure switch: BSV 06 ATEX E141 x, application area II 2 G Ex ia II B T4
  • Motor overload switch: PTB 01 ATEX 3307, application area II (2) GD

Standard equipment:

  • Control box with complete monitoring electronics incl. potential-free switching contact, e.g. for controlling the on-site exhaust air system (must be installed outside the Ex zone); cable length 5 m
  • Transparent side panes made of 5 mm toughened safety glass
  • Fold-away deflector
  • Melamine resin-coated rear wall
  • Work surface made of stainless steel (V2A/1.4301) as a seal-welded sump
  • Support frame for standing work
  • Earthing link, mounted in the profile
  • EX LED lamp 6500 K

Additional components:

  • Work surface (brushed stainless steel)
  • Earthing link mounted in the profile

Specifications

Brand
asecos · 40597-901 +2
Clear height
1,110 mm
According to standard
EN 14175-3 (5.4.4)
Extraction air connection
DN 250
Nominal power consumption
145 W
Nominal voltage
230 V
Emission sound pressure level
55 dB
Maximum current consumption
1 A
Frequency
50 Hz
ATEX marking
II 2 G Ex h IIB T4 Gb
Explosion-proof
Yes
Working height, seated
720 mm
Working height, standing
900 mm
Height with support frame for seated work
2,085 mm
Height with support frame for standing work
2,265 mm
Usable work surface depth with media duct
530 mm
Usable work surface depth without media duct
640 mm
Number of extraction air outlets
1 pc
Support frame
none
Max. load of the work surface
3,000 N/m²

On request

All options (3)

Variants & pricing

  • Model
    GAP.140.180.EX
    External dimensions W x D x H
    1800 x 750 x 1400 mm
    Internal dimensions W x D x H
    1765 x 650 x 1110 mm
    Weight
    229 kg
    Differential pressure
    135 Pa
    Volumetric flow rate (recommended)
    1,070 m³/h
    Width usable work surface
    1,690 mm
    Point load per pillar
    401 N
    40597-901On request
  • Model
    GAP.140.120.EX
    External dimensions W x D x H
    1200 x 750 x 1400 mm
    Internal dimensions W x D x H
    1165 x 650 x 1105 mm
    Weight
    177.5 kg
    Differential pressure
    103 Pa
    Volumetric flow rate (recommended)
    720 m³/h
    Width usable work surface
    1,090 mm
    Point load per pillar
    325 N
    40590-901On request
  • Model
    GAP.140.090.EX
    External dimensions W x D x H
    900 x 750 x 1400 mm
    Internal dimensions W x D x H
    865 x 650 x 1110 mm
    Weight
    163 kg
    Differential pressure
    140 Pa
    Volumetric flow rate (recommended)
    720 m³/h
    Width usable work surface
    790 mm
    Point load per pillar
    287 N
    40585-901On request