Photo 1 of 4
Rubber & plastics testing
Mooney viscometer V3
Designed with a simple, robust and dependable mechanical system that produces reliable results consistently, even in humid and remote locations. The V3 combines precise temperature control with easy-to-use software. With precise digital temperature control, live on-screen monitoring, quick result comparison and CSV export for seamless data integration, the Mooney viscometer has been used for many years in locations ranging from remote rubber plantations around the world to modern R&D laboratories.
Conforming to international standards, the Mk III is a shearing-disc viscometer in which the rubber sample is compressed pneumatically in a cylindrical chamber formed by cavities in 2 opposing dies.
The viscosity is determined by measuring the torque required to turn the rotor inside the chamber, which is heated to a set temperature. As the rotor shears the sample, the torque reaction is transmitted through a worm shaft that deflects a torsion beam. A digital dial gauge measures the beam's displacement.
Application:
- Measuring and recording the viscosity and scorch of natural, synthetic or compounded rubber.
Supplied with:
- Mooney viscometer
- Large-diameter 1.5" rotor
- Accessories and tool set
- User manual
- Calibration certificate
Calibration:
- Manufacturer's calibration certificate included as standard
Specifications
- Brand
- Wallace Instruments · WAV3/2
- Standards
- ISO 289-1, ASTM D1646
- Dimensions (H × W × D, mm)
- 810 × 510 × 460
- Weight
- 127 kg
- Max. power requirement
- 1700 VA
- LED display
- 2-line, 20-character backlit display in the control panel
- Controls
- 3 sealed switches with integral LED indicators
- Die heating
- Heating elements in the upper and lower platens, 200 W per element
- Temperature range
- 80–150 °C, ±0.1 °C
- Maximum torque
- Cut-out at 200 Mooney units
- Die closure
- By pneumatic cylinder
- Air line pressure
- Max. 80 psi or 5 bar
- Gauge (excl. V3/1)
- Range 0.0–0.5; resolution 0.0005
- Pressure controls
- Twin controls for platen closure and test routine
- Operating temperature
- 10 to 40 °C; altitude up to 2000 m
- Humidity range
- 10 to 80% RH, non-condensing
On request
Variants & pricing
Mooney viscometer V3
WAV3/2On request