Photo 1 of 6
Rubber & plastics testing
O14 ageing chamber calibration kits
offers calibration kits for regular verification of the Rapid PRI ageing chamber (O14) as part of good laboratory practice. Two kits are available: a temperature calibration kit and an airflow calibration kit. They let users make sure their O14 performs at its optimum between maintenance visits.
Temperature calibration
Temperature probes are connected to a handheld meter, inserted into chamber 3 of the ageing chamber and left for approximately 20 minutes. A reading of 140.0 ±0.1 °C on the handheld meter indicates that the chambers are maintaining temperature. On completion of the calibration test, it is important to set the probe tray aside to cool before returning it to storage.
Airflow calibration
An airflow indicator is connected to the filter inlet at the rear of the O14 to show the operating airflow rate. To comply with the standard, which requires an airflow sufficient to change the air 10 times per hour, an airflow rate of 250 ± 50 cc/min is required. A low airflow reading, even within the range, can mean that the filter needs replacing. Filter life depends heavily on the environment in which the ageing chamber operates.
Application:
- Verification readings of the O14 on a weekly basis, or as instructed according to usage levels.
Supplied with the temperature calibration kit:
- Handheld meter
- Custom temperature probe tray
- AAA batteries
- User manual
Supplied with the airflow calibration kit:
- Custom airflow meter
- User manual
Calibration:
- Manufacturer's calibration certificate included as standard
Specifications
- Brand
- Wallace Instruments · WAO14-CAL-01 +1
- Standards
- ISO 2007, ISO 2930, ASTM D3194
- Operating temperature
- 5 to 40 °C; altitude up to 2000 m
- Humidity range
- 10 to 80% RH, non-condensing
On request
Variants & pricing
- Kit Box Dimensions (mm)
- 330 (h) x 380 (w) x 190 (d)
- Kit Weight
- 1.42kg
WAO14-CAL-01On request- Kit Box Dimensions (mm)
- 175 (h) x 195 (w) x 45 (d)
- Kit Weight
- 372g
WAO14-CAL-02On request
Related products

Cutting die shank
Price on request

Data input terminal
Dimensions (H × W × D, mm): 40 × 225 × 165
Price on request

Mooney viscometer calibration kits
Rotor height gauge: 1 pc · Temperature sensor unit: 1 pc
Price on request

O14 traceability fixture
Dimensions (H × W × D, mm): 182 × 162 × 337
Price on request