Skip to main content
Saint-Tech

Rubber & plastics testing

Williams parallel plate plastimeter

A manual plastimeter offering a simple, straightforward way to measure the plasticity and recovery characteristics of silicones and unvulcanised rubber compounds to 0.01 mm.

The plastimeter has two parallel platens: a lower platen that forms the base of the instrument, and an upper platen that is raised and lowered vertically by a handle. The test consists of compressing a cylindrical sample of specified volume between the two parallel plates and measuring the compressed height after a specified period. The sample is prepared from a sheet of unvulcanised rubber about 15 mm thick. When the upper platen is lowered, a force of 49 ± 0.05 N is applied to the sample. A dial gauge reads the thickness of the test sample, which is the plasticity number.

If an oven is used, the test is normally carried out at 70 °C or 100 °C. Readings may be taken at any desired interval after the load is applied; the most frequently used intervals are 3 min and 10 min.

For the recovery test, a separate thickness gauge is required to measure the thickness of the test sample accurately. It should have a 25 mm diameter presser foot actuated by a dead-weight force of 0.8 ± 0.03 N. We recommend the digital bench thickness gauge (WAS4).

Application:

  • A simple, error-free indication of the processability of uncured rubber and silicones in terms of both plasticity and recovery. Widely used in the silicone and rubber manufacturing, research and quality control industries.

Supplied with:

  • Williams parallel plate plastimeter
  • 0.01 mm dial gauge
  • 5.000 mm slip gauge
  • User manual
  • Calibration certificate

Calibration:

  • Manufacturer's calibration certificate included as standard

Specifications

Brand
Wallace Instruments · WAP2
Standards
ASTM D926
Dimensions (H × W × D, mm)
410 × 150 × 150
Weight
11 kg
Compression force
49.00 ± 0.05 N
Sample volume
2.00 ± 0.02 cm³
Operating temperature
5 to 40 °C; altitude up to 2000 m
Humidity range
10 to 80% RH, non-condensing

On request

All options (1)

Variants & pricing

  • Williams parallel plate plastimeter

    WAP2On request